Article 1 
For the purposes of this Regulation, the following definitions shall apply:

((a)) ‘device’ means active implantable medical devices as defined in  regulation 2(1) of the Medical Devices Regulations 2002;
((b)) ‘conformity assessment body’ means a body which performs calibration, testing, certification and inspection activities under Article R1(13) in Annex I to Decision No 768/2008/EC of the European Parliament and of the Council;
((c)) “approved body” has the same meaning as in regulation 2(1) of the Medical Devices Regulations 2002;
((d)) “accreditation” means an attestation by a national accreditation body conveying formal recognition that a conformity assessment body is competent to carry out a specific activity;
((e)) ...
((f)) ...
((g)) ‘on-site assessment’ means a verification in the premises of the body or of one of its subcontractors or subsidiaries by the  Secretary of State;
((h)) ‘surveillance on-site assessment’ means a periodic routine on-site assessment which is neither the on-site assessment undertaken for the initial designation, nor the on-site assessment undertaken for the renewal of the designation;
((i)) ‘observed audit’ means  the Secretary of State’s  assessment of the performance of a notified body’s audit team in the premises of the body's client;
((j)) ‘functions’ means the tasks to be fulfilled by the body’s staff and its external experts, namely: auditing of the quality systems, product related technical documentation review, review of clinical evaluations and investigations, device testing and, for each of the previously mentioned items, the final review and decision making thereon;
((k)) ‘subcontracting’ means the transfer of tasks to one of the following:

((i)) a legal person;
((ii)) a natural person who further delegates these tasks or parts thereof;
((iii)) several natural or legal persons who jointly perform these tasks.
Article 1a 
In this Regulation, any reference to Annex 8 to Directive 90/385 or to Annex XI to Directive 93/42 is to be construed as a reference to those Annexes as they applied immediately before IP completion day and as modified by Schedule 2A to the Medical Devices Regulations 2002.
Article 2 
The criteria set out in Annex 8 to Directive 90/385/EEC or in Annex XI to Directive 93/42/EEC shall be applied as laid down in Annex I.
Article 3 

1. When applying for designation as  an approved body, a conformity assessment body shall use the application form set out in Annex II. If the conformity assessment body submits the application and documents annexed to the application on paper, it shall also submit an electronic copy of the application and its annexes.The application shall specify the conformity assessment activities, the conformity assessment procedures and the fields of competence for which the conformity assessment body wishes to be  approved, the latter by indicating the codes  found in Guidance: UK approved bodies for medical devices https://www.gov.uk/publications/medical-devices-uk-approved-bodies-for-medical-devices.
2. The Secretary of State  shall assess that body in accordance with an assessment check-list that covers at least the items listed in Annex II. The assessment shall include an on-site assessment....
3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7. ...The validity of the designation shall be limited up to a maximum of five years.
Article 4 

1. An extension of the scope of the  approved  body's designation may be granted in accordance with Article 3.
2. A designation as  an approved  body may be renewed in accordance with Article 3 before the end of the validity period of the previous designation.
3. For the purposes of paragraph 2, the procedure set out in Article 3(2) shall include, where appropriate, an observed audit.
4. Extension and renewal procedures may be combined.
5. An approved body, within the meaning of regulation A45(1)(b) of the Medical Devices Regulations 2002, whose designation does not have a stated validity period or has a validity period exceeding five years, is to be subject to a renewal within five years of IP completion day
6. By way of derogation from paragraph 2, during the period from 19 May 2020 to 25 May 2021, the designating authority of a Member State, in extraordinary circumstances resulting from the COVID-19 pandemic and due to the adoption of Regulation (EU) 2020/561 of the European Parliament and of the Council deferring the application of certain provisions of Regulation (EU) 2017/745 of the European Parliament and of the Council, may decide to renew a designation as notified body without having recourse to the procedures laid down in Article 3.In order to decide on the renewal of a designation as notified body in accordance with the first subparagraph, the designating authority shall carry out an assessment in order to verify the continuous competence of the notified body and its ability to perform the tasks for which it was designated.The decision on the renewal of a designation as notified body in accordance with this paragraph shall be adopted before the end of the validity period of the preceding designation and shall automatically become void at the latest on 26 May 2021.The designating authority shall notify the Commission of its decision, giving substantive reasons therefore, on the renewal of a designation as notified body in accordance with this paragraph by means of the ‘New Approach Notified and Designated Organisations’ Information System.The Commission may require a designating authority to provide it with the results of the assessment supporting the decision on the renewal of a designation as notified body in accordance with this paragraph, as well as the outcome of related surveillance and monitoring activities, including those referred to in Article 5.
Article 5 

1. For the purpose of surveillance, the  Secretary of State  shall assess an appropriate number of  approved  body’s reviews of the manufacturer’s clinical evaluations and shall carry out an appropriate number of file reviews, surveillance on-site assessments and observed audits at the following intervals:
(a) at least every 12 months for  approved  bodies with more than 100 clients;
(b) at least every 18 months for all other  approved  bodies.That designating authority shall, in particular, examine changes which have occurred since the last assessment and the work the  approved  body has performed since that assessment.By way of derogation from the first and second subparagraphs, in exceptional circumstances relating to the COVID-19 pandemic that temporarily prevent the designating authority of a Member State from carrying out surveillance on-site assessments or observed audits, it shall carry out any measures to ensure an adequate level of surveillance that remain possible under those circumstances in addition to the assessment of an appropriate number of the  approved  body’s reviews of the manufacturer’s technical documentation, including clinical evaluations.  The Secretary of State  shall examine changes to the organisational and general requirements set out in Annex II that have occurred since the last on-site assessment and the activities the  approved  body has performed thereafter.
2. Surveillance and monitoring conducted by the  Secretary of State  shall appropriately address subsidiaries.
3. The Secretary of State shall continuously monitor that body to ensure ongoing compliance with the applicable requirements. The Secretary of State shall provide for a systematic follow-up of complaints, vigilance reports and other information, including from outside the United Kingdom, which might indicate the non-fulfilment of the obligations by an approved body or its deviation from common or best practice.In addition to surveillance or renewal on-site assessments, the  Secretary of State  shall initiate unannounced or short-notice on-site assessments if those on-site assessments are needed to verify compliance.
Article 6 
The Secretary of State may investigate cases regarding the competence of an approved body or the fulfilment of the requirements and responsibilities to which an approved body is subject under the Medical Devices Regulations 2002.
Article 7 
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Article 8 
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Article 9 
Where designation is based on accreditation within the meaning of Regulation (EC) No 765/2008, the Secretary of State shall ensure that the accreditation body that has accredited a particular approved body is kept informed of incident reports and other information that relate to matters under the control of the approved body when the information may be relevant for the assessment of the performance of the approved body. The Secretary of State must ensure that the accreditation body in charge of the accreditation of a particular conformity assessment body is kept informed of findings relevant for the accreditation. The accreditation body shall inform the Secretary of State of its findings.
Article 10 
This Regulation shall enter into force on the twentieth day following that of its publication in the Official Journal of the European Union.
It shall apply to extension of designations as from 25 December 2013.
...
ANNEX I
1. 

1.1. The conformity assessment body shall be a third-party body that is independent of the manufacturer of the product in relation to which it performs conformity assessment activities. The conformity assessment body shall also be independent of any other economic operator having an interest in the product as well as of any competitor of the manufacturer.
1.2. That conformity assessment body shall be organised and operated so as to safeguard the independence, objectivity and impartiality of its activities. The conformity assessment body shall have procedures in place that effectively ensure identification, investigation and resolution of any case in which a conflict of interests may arise, including involvement of its staff in consultancy services in the field of medical devices prior to taking up employment with the body.
1.3. That conformity assessment body, its top management and the personnel responsible for carrying out the conformity assessment tasks shall not:

((a)) engage in any activity that may conflict with their independence of judgement or integrity in relation to conformity assessment activities for which they are notified;
((b)) offer or provide any service which may jeopardise the confidence in their independence, impartiality or objectivity. In particular, they shall not offer or provide or have offered or provided, during the last three years, consultancy services to the manufacturer, his  UK responsible person, a supplier or a commercial competitor as regards ... requirements for the design, construction, marketing or maintenance of the products or processes under assessment. This does not preclude conformity assessment activities for manufacturers and economic operators mentioned above or general training activities relating to medical device regulations or related standards that are not client specific.
1.4. The conformity assessment body’s top level management and its assessment personnel shall be impartial. The remuneration of the top level management and assessment personnel of a conformity assessment body shall not depend on the number or the results of assessments carried out.
1.5. ...
1.6. The conformity assessment body shall ensure and document that the activities of its subsidiaries or subcontractors, or of any associated body, do not affect its independence, impartiality or objectivity in its conformity assessment activities.
1.7. The requirements of points 1.1 to 1.6 do not preclude exchanges of technical information and regulatory guidance between a body and a manufacturer seeking their conformity assessment.

2. 

2.1. Subcontracting shall be limited to specific tasks. The subcontracting of the auditing of quality management systems or of products related reviews in its entirety is not allowed. The conformity assessment body shall in particular keep internal the review of the qualification and the monitoring of the performance of the external experts, the experts’ assignment to specific conformity assessment activities, and the final review and decision-making functions.
2.2. Where a conformity assessment body subcontracts specific tasks or consults external experts related to the conformity assessment, it shall have a policy describing the conditions under which subcontracting or the consultation of external experts may take place. Any subcontracting or consultation of external experts shall be properly documented and be subject to a written agreement covering, among others, confidentiality and conflict of interests.
2.3. The conformity assessment body shall establish procedures for assessing and monitoring the competence of all subcontractors and external experts used.

3. 

3.1. At all times and for each conformity assessment procedure and each kind or category of products in relation to which it has been or wishes to be notified, a conformity assessment body shall have within its organisation the following elements:

((a)) the necessary administrative, technical, clinical and scientific personnel with technical and scientific knowledge and sufficient and appropriate experience relating to medical devices and the corresponding technologies to perform the conformity assessment tasks, including the assessment of clinical data;
((b)) a documented process for the conduct of the conformity assessment procedures for which it is designated taking into account their respective specificities, including legally required consultations, in respect of the different categories of devices covered by the scope of notification, ensuring transparency and the ability of reproduction of those procedures.
3.2. The conformity assessment body shall have the necessary personnel and shall possess or have access to all equipment and facilities needed to perform properly the technical and administrative tasks entailed in the conformity assessment activities in relation to which it has been notified.
3.3. The conformity assessment body shall have at its disposal the financial resources required to conduct its conformity assessment activities and related business operations. It shall document and provide evidence of its financial capacity and its sustainable economic viability, taking into account specific circumstances during an initial start-up phase.
3.4. The conformity assessment body shall have a quality management system in place and operating.
3.5. The experience and knowledge of the personnel responsible for carrying out conformity assessment activities shall be interpreted as including the following:

((a)) sound scientific, technical and vocational training, in particular in the relevant fields of medicine, pharmacy, engineering or other relevant sciences, covering all the conformity assessment activities in relation to which the body has been notified or wishes to be notified;
((b)) substantial relevant experience covering all the conformity assessment activities in relation to which the body has been notified or wishes to be notified;
((c)) satisfactory knowledge of the requirements of the assessments they carry out and adequate authority to carry out those assessments;
((d)) appropriate knowledge and understanding of the relevant provisions of the medical devices legislation and of the applicable harmonised standards;
((e)) the ability to draw up certificates, records and reports demonstrating that assessments have been carried out.
3.6. The conformity assessment body shall establish and document qualification criteria and procedures for selection and authorisation of persons involved in conformity assessment activities (knowledge, experience and other competence required) and the required training (initial and ongoing training). The qualification criteria shall address the various functions within the conformity assessment process (e.g. auditing, product evaluation/testing, design dossier/file review, decision-making) as well as the devices, technologies and areas (e.g. biocompatibility, sterilisation, tissues and cells of animal origin, clinical evaluation) covered by the scope of designation.
3.7. The conformity assessment body shall have procedures in place to ensure that its subsidiaries operate on the basis of the same operating procedures and with the same stringency as its headquarters.
3.8. Where subcontractors or external experts are used in the context of the conformity assessment, in particular regarding novel, invasive and implantable medical devices or technologies, the conformity assessment body shall have adequate internal competence in each product area for which it is designated to direct the conformity assessment, to verify the appropriateness and validity of expert opinions and make the decision on the certification. The internal competence requested shall cover technological, clinical and auditing aspects.

4. 

4.1. The conformity assessment body shall take out appropriate liability insurance corresponding to the conformity assessment activities for which it is notified, including the possible suspension, restriction or withdrawal of certificates, and the geographic scope of its activities, unless liability is assumed by the State under domestic legislation or the Member State itself carries out the inspections directly.

5. 

5.1. The conformity assessment body shall ensure that confidentiality of the information which comes into its possession during the performance of the conformity assessment activities is observed by its personnel, committees, subsidiaries, subcontractors or any associated body, except when disclosure is required by law. To this end, it shall have documented procedures in place.
5.2. The personnel of a conformity assessment body shall observe professional secrecy with regard to all information obtained in carrying out their tasks, except in relation to the designating authorities and the competent authorities or the Commission. Proprietary rights shall be protected. To this end, the conformity assessment body shall have documented procedures in place.

ANNEX II

Designating authority: …
Name of the applying conformity assessment body: …
Previous name (if applicable): …
Approved  Body number (if applicable): …
Address: …
…
…
…
Contact person: …
E-mail: …
Telephone: …
Legal form of the conformity assessment body: …
Company registration number: …
At company register: …
…
…
The following documents shall be added. In case of extension or renewal, only new or modified documents shall be submitted.

Item/issue Corresponding Annex I section Attachment number + Reference(Section/page)
ORGANISATIONAL AND GENERAL REQUIREMENTS
Legal status and organisational structure
1 Company statutes  
2 Extract of company registration or enrolment act (Company register)  
3 Documentation on the activities of the organisation to which the conformity assessment body belongs (if any) and its relationship with the conformity assessment body  
4 Documentation on entities the conformity assessment body owns (if any) ...  
5 Description of legal ownership and the legal or natural persons exercising control of the conformity assessment body  
6 Description of organisational structure and the operational management of the conformity assessment body  
7 Descriptions of functions, responsibilities and authorities of top-level management  
8 List of all staff who have an influence in the conformity assessment activities  
9 Documentation on other services provided by the conformity assessment body (if any) (e.g. consultancy relevant to devices, training etc.)  
10 Documentation on accreditation(s) relevant to this application  
Independence and impartiality
11 Documentation on structures, policies and procedures to safeguard and promote the principles of impartiality throughout the organisation, personnel and assessment activities, including ethical rules or codes of conduct  
12 Description of how the conformity assessment body ensures that the activities of subsidiaries, subcontractors and external experts do not affect its independence, impartiality or objectivity  
13 Documentation on the impartiality of the top-level management and personnel involved in conformity assessment activities, including their remuneration and bonuses  
14 Documentation on conflict of interest and resolution of potential conflict procedure/form  
15 Description of independence of the conformity assessment body from  the Secretary of State  
Confidentiality
16 Documentation on professional secrecy procedure including protection of proprietary data  
Liability
17 Documentation of the liability insurance, proof that the liability insurance covers cases where the  approved body  may be obliged to withdraw or suspend certificates  
Financial resources
18 Documentation of the financial resources required to conduct the conformity assessment activities, related operations, including the ongoing commitments for certificates issued to demonstrate the continuing viability of the  approved body  and consistency with the range of products certified  
Quality system
19 Quality Manual and a list of related documentation on the implementation, maintenance and operation of a quality management system, including policies for assignment of personnel to activities and their responsibilities  
20 Documentation on the procedure(s) for control of documents  
21 Documentation on the procedure(s) for control of records  
22 Documentation on the procedure(s) for management review  
23 Documentation on the procedure(s) for internal audits  
24 Documentation on the procedure(s) for corrective and preventive actions  
25 Documentation on the procedure(s) for complaints and appeals  
Resource requirements
General
26 Description of own laboratories and testing facilities  
27 Employment contracts and other agreements with internal personnel, in particular in relation to impartiality, independence, conflict of interest (attach a standard contract template)  
28 Contracts and other agreements with subcontractors and external experts, in particular for impartiality, independence, conflict of interest (attach a standard contract template)  
Qualification and authorisation of personnel
29 List of all permanent and temporary personnel (technical, administrative etc.) including information on professional qualification, past experience and the types of contracts held  
30 List of all external personnel (e.g. external experts, external auditors) including information on professional qualification, past experience and on the types of contracts held  
31 Qualification matrix linking the body's staff and its external experts to the functions to be accomplished by them and to the fields of competence for which the body has been  approved  or wishes to be  approved  
32 Qualification criteria for the different functions (see point 31)  
33 Documentation on the procedure(s) for selection and assignment of internal or external personnel involved in the conformity assessment activities, including conditions for the attribution of tasks to external personnel and the supervision of their expertise  
34 Documentation demonstrating that the management of the conformity assessment body has appropriate knowledge to set up and operate a system for:
— the selection of the personnel deployed during the conformity assessment,
— the verification of the knowledge and experience of this personnel,
— the assignment of the personnel to their tasks,
— the verification of the performance of the personnel,
— the definition and the verification of their initial and ongoing training  
35 Documentation on the procedure assuring ongoing monitoring of competences and performance monitoring  
36 Documentation on standard training programmes conducted by the conformity assessment body relevant to the conformity assessment activities  
Subcontractors
37 List of all subcontractors (not individual external experts) used for conformity assessment activities  
38 Subcontractor policy and procedure  
39 Documentation demonstrating adequate core competence within the conformity assessment body to assess, select, contract, and to verify the appropriateness and validity of subcontractor activities  
40 Examples of standard template contract, prohibiting further subcontracting by legal persons and specifically including provisions to ensure confidentiality and conflict of interest management with subcontractors (attach examples)  
Process
41 Documentation on procedures relating to conformity assessment activities and other related documents reflecting the scope of conformity assessment activities including, in particular procedures relating to:
— Qualification and classification
— Quality system assessments
— Risk management
— Pre-clinical data evaluation
— Clinical evaluation
— Representative sampling of technical documentation
— Post-market clinical follow up
— Communications from the Secretary of State or other regulatory authorities
— Communication and analysis of the impact of vigilance reports on device certification
— Consultation procedures for drug-device combination products, devices utilising animal tissue, devices utilising human blood derivatives
— Review and decision making on certificate issuance including approval responsibilities
— Review and decision making on certificate suspension, restriction, withdrawal and refusal including approval responsibilities  
42 Checklists, templates, reports and certificates used for the conformity assessment activities  

Name and signature of an authorised representative of the applicant conformity assessment body (unless electronic signature is accepted)  Place and date